A bump test exposes the sensors to challenge gas long enough to confirm that gas reaches the sensors and that the instrument activates its alarms. It is a functional check, not a measurement of accuracy. Portable monitors should be bump-tested or calibration-checked before each day’s use when required by the manufacturer’s instructions and site procedure.